15000-MNL-01 Rev A
31
the error persists, please contact a
technical representative.
T_COM_HAL_COM_ERR
-20035
HAL/low level
communications failure.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
FIRMWARE_SYSTEM_IN_WARMUP
-20100
Attempted to go to RUN
mode when system still in
WARMUP
Wait for the system to exit WARM-UP
mode and transition to RUN mode.
FIRMWARE_MAX_SHUTTERS_OPEN
-20103
Attempted to open a
shutter when maximum
number already open
Only 4 shutters can be open at one time.
If this fault persists while less than four
shutters are open, unplug the unit and
wait 10 seconds before reapplying power
to the system.
FIRMWARE_NECSEL_CONTROLLER_TEMP_SERVO
-20259
Necsel Controller
temperature servo error
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
FIRMWARE_NECSEL_CONTROLLER_LASER_UNDER_TEMP -20262
Necsel Controller laser
under temp error
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
FIRMWARE_NECSEL_CONTROLLER_LASER_OVER_TEMP
-20263
Necsel Controller laser
over temp error
Close all shutter and let the system cool
down. If the error persists, verify that the
unit is operating within the specified
ambient temperature range and nothing
is blocking the air inlet or exhaust.
FIRMWARE_NECSEL_CONTROLLER_CURRENT_AT_RAIL
-20265
Necsel Controller current
at rail error
Please contact a technical representative
if this error persists.
PC_APP_WRITE_COMMAND_ERR
-40006
LDI UI - communication
error with LDI
Make sure the LDI is powered on. Check
USB cable connection to the host PC and
the LDI. Disconnect the LDI from the PC,