15000-MNL-01 Rev A
29
SYSTEM_FAULT_COVER_REMOVED
Interlock - Cover
Removed.
Please contact a technical representative
if this error persists.
SYSTEM_FAULT_INTERLOCK_USER_OPEN
Interlock - Remote Open.
Verify that the eyepiece interlock on the
back of the LDI is properly installed and in
the proper state.
SYSTEM_FAULT_INTERLOCK_FIBER1_OPEN
Interlock - Novalum Fiber
Open.
Verify that the interlock switch in the
RGGBBV fiber routing slot on the front of
the LDI is completely depressed and the
"Laser Ready" light is blue. See Figure 11.
SYSTEM_FAULT_INTERLOCK_FIBER2_OPEN
Interlock - 555 Fiber Open. Verify that the interlock switch in the 555
fiber routing slot on the front of the LDI is
completely depressed and the "Laser
Ready" light is blue. See Figure 11.
SYSTEM_FAULT_OVER_CURRENT
Over current.
Please contact a technical representative
if this error persists.
SYSTEM_FAULT_NECSEL_OVER_TEMP
Necsel Over Temp.
Close all shutters and let the system cool
down. If the error persists, verify that the
unit is operating within the specified
ambient temperature range and nothing
is blocking the air inlet or exhaust.
SYSTEM_FAULT_NECSEL_UNDER_TEMP
Necsel Under Temp.
Verify that the unit is operating within
the specified ambient temperature
range. Please contact a technical
representative if this error persists.
SYSTEM_FAULT_NOVALUM_OVER_TEMP
Novalum Over Temp.
Close all shutters and let the system cool
down. If the error persists, verify that the
unit is operating within the specified
ambient temperature range and nothing
is blocking the air inlet or exhaust.
SYSTEM_FAULT_NOVALUM_UNDER_TEMP
Novalum Under Temp.
Verify that the unit is operating within
the specified ambient temperature