34
35
2�
To create or edit a custom cycle, select either
New
, or
Open
an existing custom cycle (Figure 38)
X
3M
™
Cycle Programmer
Actions
Cycle Name:
GS5X
GS8X
Operator
Supervisor
Help
Access Level
Model
Name
File
Commands
One Column
Two Columns
Columns
?
Estimated total cycle time
Preheat
Air removal
Chamber test Conditioning EO injection EO exposure EO removal
Flushing
Aeration
15 Hours, 13 Minutes
9. Aeration
55
12.0
34
0
60
Maximum
Set Point
Minimum
Parameter
Aeration temperature
Minimum aeration time
Units
999
˚C
hr
6. EO exposure
60
0
999
Maximum
Set Point
Minimum
Parameter
Duration
Units
min
1. Preheat
55
60
0
34
30
0
60
Maximum
Set Point
Minimum
Parameter
Cycle temperature
Timeout
Hold time
Units
200
999
˚C
min
min
2. Air removal
95
160
100
95
100
25
99
Maximum
Set Point
Minimum
Parameter
Vaporizer temperature
Vacuum level
Vacuum rate
Units
160
100
˚C
mbar
%
3. Chamber test
6
5
99
Maximum
Set Point
Minimum
Parameter
Duration
Units
min
4. Conditioning
30
90
Maximum
Set Point
Minimum
Parameter
Relative Humidity (RH)
Timeout
Units
200
min
%
Hold time
0
999
%
RH injection rate
Slow
Fast
RH lower limit
10
50
%
RH upper limit
70
90
%
60
50
30
Fast
10
90
No RH
Injection
5. EO injection
No
100
Yes
Maximum
Set Point
Minimum
Parameter
Simulate EO
Vacuum level
Units
160
mbar
Minimum pressure increase
100
550
mbar
RH lower limit
0
50
RH upper limit
70
90
%
%
No
160
200
40
80
7. EO removal
100
160
25
100
100
Maximum
Set Point
Minimum
Parameter
Vacuum rate
Vacuum level
Units
160
8. Flushing
5
2
99
Maximum
Set Point
Minimum
Parameter
Number of flushing cycles
Units
25
Vent rate
100
%
Hold time: Top
4
300
0
90
180
1
99
min
mbar
Vacuum level
100
100
300
Hold time: Bottom
0
99
min
Locked aeration time
0
999
min
No exhaust hood aeration time
0
999
min
25
Vacuum rate
100
%
100
%
mbar
Figure 38
3M™ Cycle Programmer