74
3M
™
Steri‑Vac
™
Sterilizer / Aerator GSX Series
– Operator’s Manual
13.9. Accessing the Chamber – Aeration Not Complete
During user programmed aeration, the 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series door can only be opened with a Supervisor PIN (Figures 90 and 91) Aeration timing will be interrupted if
the door is opened Once the door is closed and aeration temperature is reached, aeration timing will continue
The vent hood (ie exhaust hood) is monitored for appropriate air flow in standard cubic feet per minute (SCFM) If the GXS Series sterilizer detects the air flow is too low, the sterilizer door will
remain locked until a minimum of three hours of aeration is fulfilled
1�
To access the chamber when aeration is not complete, press the
Open Door
button on the touch screen On double door units, press the
Open Unload Door
button
2�
Enter the Supervisor or Service PIN
3�
Close the GSX Series sterilizer door to continue aeration
Sterilizer
Aeration time elapsed (hh:mm):
0:05
1:00
11:55
55.0
˚
C
04/20/2015
20:56:12
Gas expose time elapsed (hh:mm):
Time remaining (hh:mm):
Cycle:
Stage:
Count:
33
QA 38 Fast
Aeration paused
84.5
kPa
Cancel
Cancel
Enter supervisor PIN:
1
4
7
2
5
8
0
3
6
9
OK
OK
Figure 90
Supervisor PIN Required to
Pause Aeration
Stop
Sterilizer
Aeration time elapsed (hh:mm):
0:05
1:00
11:55
55.0
Menu
Menu
˚
C
04/20/2015
20:54:20
Gas expose time elapsed (hh:mm):
Cycle time remaining (hh:mm):
Cycle:
Stage:
Count:
33
Aeration paused
Open door
Show chart
98.0
kPa
GSX 55C
Figure 91
Aeration Stage Paused