24
25
Table 3 outlines the parameters in
Site Setup>Preferences
including description, default value, and minimum access level
Parameter Display
Name
Parameter Description
Data Range
(or Format)
Default
Value
Minimum Access
Level to Edit
Audible tone on cycle
failure
When enabled, system will produce an audible tone upon cycle failure The audible
notification will sound at a pulsed rate of two (2) seconds on, one (1) second off, for
total of 30 seconds
Disabled, Enabled
Disabled
Supervisor
Audible tone on cycle
complete
When enabled, system will produce an audible tone upon cycle completion The
audible notification will sound continuous for total of 20 seconds
Disabled, Enabled
Disabled
Supervisor
Print on cycle complete
When enabled, system will automatically print cycle reports on sterilizer printer
Disabled, Enabled
Enabled
Supervisor
Cycle report preference
Allows user to select default cycle report type for printing and saving
Graph, Table, or
Detailed
Graph
Supervisor
Sensor to plot
Allows user to select default sensors that will be plotted on graph of cycle reports
Control, Monitor, Both
Control
Supervisor
Screen timeout
Allows user to select amount of time in minutes that must pass without a detected
touch before the system will dim the display
[1:60] minutes
10 minutes
Supervisor
Idle mode chamber
temperature
When enabled, the system will maintain user‑entered target temperature while in Idle
mode
Off,
GSX [30 ‑ 60] °C
Off
Supervisor
Pressure unit
Allows user to select preferred units for pressure on display and select reports
mbar,
kPa
mbar
Supervisor
User cycle count
Will count cycles performed between supervisor access level resets Incremented at
start of a cycle
[0 ‑ 2147483647]
0
Supervisor‑ Reset to
zero only
Reset (User cycle count)
When enabled with supervisor level access, will reset User cycle count indicator
Disabled, Enabled
0
Supervisor‑ Reset to
zero only
Abated aeration time
If the system is configured to use an Abator, this control will allow the user to enter
how many hours into aeration the system shall request the Abator to run Abator start
will remain on for this amount of time after the start of aeration
0 to 999:30 hours,
adjustable by 05 hour
increments
0
Supervisor
Table 3 Site Setup Preferences Parameters
The
Site Setup>Cycle
tab provides the option to select the cycles that will appear on the Main Screen for the Operator, Supervisor, and Service Figure 22 illustrates the options when selecting
the Operator Cycles for the Main Screen
Sterilizer
22.2
˚
C
04/20/2015
12:37:05
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
Return
Apply changes
Cycle access level
Touch check box to select cycles listed on start screen.
Touch row to change values.
Operator
Cycle
GSX 38C
GSX 55C
36.0 hrs
12.0 hrs
Minimum aeration
Setup
Preferences Cycle Setup Date-Time Network Site
Figure 22
Setup Cycle Options for
Operator Cycles