30
Basic setups
Measurement setup (for community noise)
Models SE and DL
C-A measurement settings
The C-A measurement is determined from concurrent measurements of the
SPL by both meters. It can only be made if certain settings for Meter 1 and
Meter 2 are compatible. If any incompatibility exists, the L
c-a
selection in this
screen will be marked N/A (Not Available).
If you want to activate L
c-a
, have the following settings in place before you
visit the Measures screen.
•
Thresholds
- Both meters set to “Off” or the same value.
•
Exchange rates
- Both meters set to the same value.
•
Time response
- Both meters set to the same value.
•
Frequency weightings
- Meter 1 must be set to A-weighting; Meter 2
must be set to C-weighting.
Figure 3-6 is an example of the appropriate C-A settings for Meter 1 (left
side) and Meter 2 (right side). (Please see “Meter 1/Meter 2 Parameters” on
page 27 for setup details.)
Figure 3-6: C-A measurement settings for Meter 1 and Meter 2
Meter 1
•
Press softkey to
toggle between
Meter 1/Meter 2
Slow response A-weighting
Slow response C-weighting Meter 2
Summary of Contents for SP-DL-1
Page 1: ...3M Personal Safety Division 3MTM SoundPro SE DL SeriesSound Level Meters SoundProUser Manual ...
Page 13: ...xii Table of Contents Models SE and DL This page left blank intentionally ...
Page 35: ...22 Turning off Navigating Models SE and DL This page left blank intentionally ...
Page 69: ...56 Communications GPS Models SE and DL This page left blank intentionally ...
Page 89: ...76 Back erasing Stopping a session Models SE and DL This page left blank intentionally ...
Page 137: ...AC DC output DC output 124 Models SE and DL This page left blank intentionally ...
Page 149: ...Appendix A 136 Models SE and DL Block Diagram for SoundPro ...
Page 161: ...Appendix B 148 Models SE and DL This page left blank intentionally ...