Pannoramic SCAN 150 1.15 User’s Guide
5 Troubleshooting and Maintenance
5.2.1 Fluorescent Applications (optional)
The shutter of the HXP 120 cannot be triggered, in spite of trigger connection through AxioCam
MRm Rev3.
•
HXP 120: Shutter is permanently open
•
Activate shutter remote control with the button at the front of the HXP 120.
The switching of the Lumencor cannot be triggered, in spite of trigger connection through the
camera.
•
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
•
HXP 120: Shutter is closed
•
The opening of the iris contained in the lamp module for fluorescent applications
can be varied by means of the wheel located at the front side of the module. Open
the iris completely.
•
Intensity is too low
•
Increase intensity with the rotary button at the front of the HXP 120.
•
HXP 120 is switched off
•
Switch light source on.
•
When using Lumencor
•
Contact 3DHISTECH Service for support.
March 13, 2012 - Rev. 1