Pannoramic SCAN 150 1.15 User’s Guide
4 Software Description
•
Do not use multilayer mode on empty field-of-views –
If selected, the
extended focus or the Z-stack function will not run on the empty (not
containing specimen) field of views.
•
Use stitching –
By selecting, this function enables seamless stitching of digitized
tiles.
Focus Exposure Time
This function is used for defining time length used for determination of focus but is inactive
when using AxioCam MRm Rev.3 and pco.edge cameras.
Exposure Time
•
Calculate exposure time for each area
– Calculates exposure time that is specific to each
area, and that varies according to the different characteristics.
•
Calculate exposure time for full slide
– After analyzing the slide, the program uses the same
smallest exposition time for each area to omit over-exposition.
Note
:
This option is used only if no exposure times were manually
predefined. It is recommended to use the second option.
Multi Filter Change Mode
This is for a multi-channel digitization. To optimize speed, filter change should take place
after the digitization of the slide with one channel. Filters should be changed for each field
of view if very precise colocalization is required.
•
Change filters at each area
•
Change filters at each field of view
Fluorescent Swap File Location
This determines where the software stores the swap files necessary for scanning.
2012. March 13.
3DHISTECH Ltd.