Pannoramic MIDI 1.16 User’s Guide
4 Software Description
4.5.3 Preview tab
The Preview tab offers options for threshold setting and selection of the area to be scanned.
Figure 48 – Preview tab
Use threshold
•
Scan specimen with auto threshold –
This automatically detects the specimen in the
preview image.
•
Scan specimen with user set threshold –
This sets the sensitivity of the specimen detection.
To set the desired threshold value manually, change the setting in the
Specimen threshold
level
selection box in the range of 0-255. The default value is 35.
Setting smaller values result in a more sensitive scanning, meaning that more areas will be
digitized, however, this may also include specks or other imperfections of the slide.
Greater values result in a less sensitive scanning, but it is possible that some of the smaller
specimen areas will not be digitized.
January 16, 2013 - Rev 1
3DHISTECH Ltd.
79(127)