Pannoramic MIDI 1.16 User’s Guide
4 Software Description
•
Do not use multilayer mode on empty field-of-views –
If selected, the
extended focus or Z-stack function will not run on empty (not containing
specimen) field of views.
•
Use stitching –
By selecting, this function enables seamless stitching of digitized
tiles.
Focus Exposure Time
This function is used for defining time length used for determination of focus but is inactive
when using AxioCam MRm Rev.3 and pco.edge cameras.
Image corrections
Flat field correction
If this function is active the scanned slide compensates itself with a special method (light
intensity statistical data, not to modify the average intensity of the slide. Quantitative
measurements can be made on the slide.
•
Sharpening
– image contrast is increased (the image may become noisier)
Note
:
Do not use this option if you want to take quantitative measurements,
since it may minimally affect the results.
Exposure Time
•
Calculate exposure time for each area
– Calculates exposure time that is specific to each
area, and that varies according to the different characteristics.
•
Calculate exposure time for full slide
– After analyzing the slide, the program uses the same
smallest exposition time for each area to omit over-exposition.
Note
:
This option is used only if no exposure times were manually predefined,
and it is recommended to use the second option.
January 16, 2013 - Rev 1
3DHISTECH Ltd.
77(127)