Pannoramic MIDI 1.16 User’s Guide
5 Troubleshooting and Maintenance
9.
Screw the lateral cover-plate down to the housing.
10.
Connect power plug.
11.
Set hour meter (time counter) to zero.
5.3.4 Removing a slide
If a slide got jammed or has suffered breakage during the process of digitization, stop the process
and, if possible, move the slide holder to a position where the slide can be easily removed by clicking
the
Emergency Manual Slide Removal Position
button (see section
4.3.6 Service tab
).
Warning!
Crush hazard! Before you remove a jammed slide, cancel
the running
digitization process, terminate the software session and turn the power
off.
Any manipulation performed on a mechanical component must be followed by a complete device
initialization routine (software restart).
Figure 75 – Removing a slide
1.
Push protective covers apart (see section
5.3.1 Removing the protective covers
).
2.
Position the slide
(2)
.
3.
Remove the slide carefully.
January 16, 2013 - Rev. 1
3DHISTECH Ltd.
118
(