Pannoramic MIDI 1.16 User’s Guide
5 Troubleshooting and Maintenance
Warning!
Hot surface! Wait at least twenty minutes to allow the lamp surface to
cool down sufficiently.
1.
Disconnect power plug (separate all poles from the mains).
2.
Unfix screws of lateral cover plate and remove it from the unit housing (see
Figure 71
).
Figure 71 – Removing the lateral cover of the HXP 120
3.
Press plug securing latches inward (
1
) then disconnect plug of the lamp module (
2
) (see
Figure 72
).
Figure 72 – Disconnect plug
January 16, 2013 - Rev. 1
3DHISTECH Ltd.
116
(