E. Settings
Model AQSync Air Quality Monitoring Station Manual Rev. A-1
30
E.4
CO Module Settings
There are 2 identical CO sensors in most versions of
the AQSync, labeled as “A” and
“B” in the CO calibration screen. The measured value of CO shown on the touch screen
and found in the data stream is the average of the 2 sensors. Calibration values for
slopes and offsets can be entered here. The CO calibrations are carried out at the
factory for your AQSync, and new values would only be needed if a new calibration is
carried out by the user or during factory servicing.
E.5
Weather Transmitter Settings
The offsets for temperature, pressure, humidity,
and wind direction can be specified using the
Weather Transmitter Settings screen. In the
AQSync, the factory settings are 0 for all offsets.
The following parameters of the Weather
Transmitter are pre-set when the AQSync is
shipped:
Output Interval
5 sec
Compass
Off
Wind speed
Units
meters/sec
Output format
Cartesian
Multiplier
1.0000
Threshold
NA
Direction offset
0 deg
Damping factor
0
Sample count
100
Temperature
Units
°C
Offset
0
Pressure
Units
hPa
Offset
0
Relative Humidity
Units
%
Offset
0