H. Maintaining and Troubleshooting the AQSync
Model AQSync Air Quality Monitoring Station Manual Rev. A-1
53
H.2.6
CO Module Troubleshooting
In most configurations, the AQSync has two CO sensors. For improved accuracy, the
AQSync gives an average of their measurements. The output of the two sensors can
be monitored through the CO Parameters screen. If the readings are systematically
changing, with one sensor drifting higher or lower than the other, replacement of the
sensor is indicated. Ideally, the two sensor readings should be within ~20% of each
other.
H.2.4
Service through 2B Technologies
2B Technologies offers reasonably priced customer service for instrument repairs. The
calibration service for the ozone monitor includes cleaning of the entire flow path with
methanol, testing of all components for proper function, installation of a new internal
ozone measurement scrubber and calibration against a NIST-traceable standard. 2B
Technologies also offers calibration and service for the sensors. The best way to contact
us for service is to log a customer service ticket at
. Normally, you will hear back from us by email within a few hours. Or, call
us at +1(303)273-0559.
A great deal of technical information about our instruments is posted as technical notes on
the 2B Tech website at
https://www.twobtech.com/docs/docs_tech_notes.htm
. Manuals,
brochures, software, cleaning procedures and scientific papers may be downloaded at
https://www.twobtech.com/downloads.html
. Replacement parts may be purchased by
calling us at +1(303) 273-0559 or by emailing us at