15
Initial oven setup
Page
Enter
Setup Oven
450C
Change idle temp.
Change night temp.
Customize calibration
Vacuum pump operation
Set vacuum level
Diagnostic tests
Customize country
Adjust display
Audio beeps
Set preview mode
Esc
Users can enter their individual requirments here.
With the +/-keys select in the main menu “Setup
Oven” and confirm by pressing the green “Enter”-
key.
1.3.1 Change idle temperature
This is the temperature reached by the furnace fol-
lowing the conclusion of a firing program.
This temperature should be based on the lowest
starting temperature for the individual firing cycles.
1.3.2 Change night temperature
This value displays the temperature maintained by
the furnace during night mode (see section 1.5)
1.3.3 Customize calibration
This allows the user to adjust the calibration within
certain temperature ranges. This feature changes all
the final temperatures of programs in the respecti-
ve temperature range below/above 800°C). 1472°F
1.3.4 Vacuum pump
This section determines whether the pump is main-
tained in continuous operation, or is switched off
when the required vacuum level has been reached.
It is recommended to run the vacuum pump in
the "cyclical" mode in order to increase its life.
1.3.5 Vacuum level
In this section, you can set the vacuum level for the
programs. (Run a vacuum test to check for the
allowable vaccum level ???) (see 1.3.6 diagnostic
tests, point 5 service tests).
“Load Programs from USB ”
■
Select “Load Programs from USB” using the
+ / -keys or the dialer and confirm by pressing
the green "Enter" key.
■
Select the program file you want to load from the
USB, using the + / -keys or the dialer and confirm
by pressing the green "Enter" key. When the
transfer is complete the display will return to the
“Edit Programs” menu.
1 Operation
1.3 Setup Oven
Page
Содержание VARIO 200
Страница 1: ...www zubler de VARIO 2 20 00 0 Operating instructions 01 2014 en Ceramic Furnace...