QUICK START

 

GDP-32

24

 

Section 4, Page 9 

September 2015

 

4.5

 

TIMING CARD SYNCHRONIZATION 

Synchronization of the timing card should be performed when synchronization with GPS, or an 
XMT, is required.  Every time the GDP-32

24

 is powered up, the SYNC program should be run to 

establish GPS or XMT synchronization.  If only the operating system was shut down by pressing 

 or the operating system restarted by pressing the 

 key, the SYNC program should 

be run to re-establish GPS syncing.  For manual syncing with an XMT, SYNC does not need to 
be re-run as long as Crystal Power was not shut down and battery power did not fail.   

To start the SYNC program, press 

 from the Zonge Logo screen.  Prior to synchronization 

the receiver should be connected to either a GPS antenna or to an XMT transmitter.  See 

Section 

6,  Receiver  Setup

  for  more  information  about  making  the  connections  to  a  transmitter 

controller.   

 

<0405a_24SYNC_Sats> 

4.6

 

GPS SYNCHRONIZATION 

For GPS synchronization, start the program and wait until the signals from at least 4 satellites are 
received (display line 4,

 Sats:

).  The Crystal Percent Stable should be at least 50%.  Press  

 to move to the Use GPS Syncing field and then press 

 to select Yes.  After a few 

moments the Timing card will be synchronized.   

Note:    Some  XMT-G  models  have  GPS  antennas  built  into  the  lid  of  the  case,  as  well  as  an 
external GPS antenna connector.  Use of an external antenna is preferable.  

WARNING

: Only 

one antenna should be connected at any time.  Connecting both antennas will cause a phase shift 
of the transmitter control signal.   

ON/OFF

Pg Dn

NEXT FIELD

Home

SELECT UP

Содержание GDP-32

Страница 1: ...ANUAL SYNCHRONIZATION 10 4 8 SET TIME 11 4 9 SYNC CHK 13 4 10 FIELD SURVEY PROGRAM 14 4 11 JOB_INFO SCREEN 15 4 12 LINE SETUP 16 4 13 ACQUISITION CONFIGURATION 17 4 14 CHANNEL TABLE 18 4 15 CHECK FIEL...

Страница 2: ...ck Field connections Acquire Data Check Data 6 Transfer Data Simple menus make operating the GDP 3224 straightforward The easiest way to become familiar with the receiver operation is to try the vario...

Страница 3: ...r 2015 CURSOR POSITION To position the cursor on the LCD display use the Cursor Control Keys or and To cycle through available options labels and numbers use the Rotary Entry Keys and Pg Dn NEXT FIELD...

Страница 4: ...ith an XMT transmitter controller both the receiver and XMT should be powered on for at least an hour before syncing and acquiring data Press the button located on the Front Panel The unit beeps and t...

Страница 5: ...sure all the channels to be calibrated are turned on 3 Set the frequency to the lowest frequency that will be used during the survey 4 Select the notch filter to be used for the survey Note It is good...

Страница 6: ...ecially as the environment changes will provide the best results Refer to Section 6 Setup and the sections for specific programs for more information about calibration and when to use METHOD 1SHOTCAL...

Страница 7: ...closely to confirm and identify and real problem See the System Check section of this manual for more information FREQ Set to the lowest frequency to be used for the job CYCLES At least 2 cycles for t...

Страница 8: ...GDP 3224 INSTRUCTION MANUAL September 2015 Section 4 Page 8 GDP 3224 0404a_24SYSCHK_Quick For more detailed instructions see Section 8 System Check...

Страница 9: ...zation the receiver should be connected to either a GPS antenna or to an XMT transmitter See Section 6 Receiver Setup for more information about making the connections to a transmitter controller 0405...

Страница 10: ...the receiver and XMT should be powered on for at least an hour before running SYNC If a fast beep sounds while pressing any of the crystal adjust keys no more adjustment is available on that particul...

Страница 11: ...the and keys to make adjustments After making adjustments move the cursor back to the MODE field and make sure the GPS Local Time appears correct If not adjust the Local Time Differential setting When...

Страница 12: ...GDP 3224 INSTRUCTION MANUAL September 2015 Section 4 Page 12 GDP 3224 0408b_24SYNC_TimeManl...

Страница 13: ...the internal reference Set channel 2 comp field to XMT and connect the channel inputs to the XMT Sync Out ports Turn Off all other channels not connected to a synchronized signal Press 0409a_24SNCCHK...

Страница 14: ...eters can be entered directly on the Main Menu screen Various submenu screens are accessed directly from the Main menu for entering more specific job or special setup parameters Functions such as Cali...

Страница 15: ...job information submenu where Operator Transmitter and other job based data are entered Programs such as TEM also use this screen to enter loop antenna configurations delay times and other data which...

Страница 16: ...all programs contain all the fields listed below some programs contain additional fields but all the fields in this portion of the screen should be inspected and modified if necessary TIME SERIES Set...

Страница 17: ...if necessary FREQ Select the frequency to analyze CYCL Set the number of cycles to average STACKS Set the number of repetitions of data acquisition to perform Each stack will be recorded as its own b...

Страница 18: ...l parameters Some examples are Note Refer to the appropriate Survey Program chapters for more information on setting up channels Any designator other than OFF will turn on channels for all programs wi...

Страница 19: ...ELD CONNECTIONS The Channel Table is also used to verify electrode and coil field connections prior to data acquisition Pressing starts the Contact Resistance CRES function to measure the resistance o...

Страница 20: ...to evaluate the process When data has been acquired the final results will be calculated and displayed When taking a single stack of data the GDP will display the message Press ESCAPE to discard CONT...

Страница 21: ...mode Each program has its own cache viewing features and procedures Refer to the appropriate field survey manual for more information Press the function key to return to Data Acquisition mode 0417a_2...

Страница 22: ...screen press to start the connection 0418a_GDPNET_ManualStart Enter an IP address consistent with the address of the PC or laptop The first 3 octets for the receiver address should be set to match tho...

Страница 23: ...er to the Local Area Network hub with a standard cable From the Zonge Logo screen press to start the connection Press to select the AUTOMATIC settings 0418c_GDPNET_NetStart Press to initiate the conne...

Страница 24: ...attributes and properties from the PC or laptop computer s File Explorer This can make the GDP system unstable File attributes on the GDP must NEVER be set to Read Only or Hidden FILE TRANSFER Files c...

Отзывы: