Specifications
A-14
www.zoll.com
9650-0806-01 Rev. D
Monitor/Display
Input
: 3-lead, 5-lead, or 12-lead patient cable, paddles, Multifunction or OneStep electrodes.
Type
: Color LCD, 640 x 480 pixels, 800 MCD
Sweep Speed
: 25 mm / sec or 50 mm / sec (User Selectable)
Lead Selections
: Paddles (Pads), I, II, III, AVR, AVL, AVF, V1-6.
Frequency Response
:
Pads/Paddles:
0.67 to 20Hz Limited response
3/5/12 Lead Continuous Monitoring (user selectable):
0.67 to 20Hz Limited response
0.67 to 40Hz Monitor response
0.05 to 40Hz EMS Diagnostic response
Acquired 12-lead snapshots (supervisor selectable):
0.25 to 40Hz Filtered Diagnostic response
0.05 to 150Hz Diagnostic response
Per methods a, b, c of EC11 3.2.7.2
(Automatically sets chart recorder response)
Common Mode Rejection
:
Complies with AAMI EC13-2002 section 4.2.9.10.
Tall T-Wave Rejection
:
Meets AAMI EC13-2002, section 4.1.2.1c for 0.9 mV
T-wave (0.8 mV with diagnostic response) and 1mV QRS.
Diagnostic Signals Applied to Patient Connections
:
Leads off / active noise suppression sensing circuit is < 0.1 microamps DC. The impedance
/respiration detector signal frequency is 72 ± 7 kHz at 77 microamps RMS
pseudo-sinewave into 100 ohms.
Heart Rate Range
: 30 to 300 BPM.
Heart Rate Accuracy
: +/- 3% or +/- 3BPM, whichever is greater.
Displayed Heart Rate:
Average of last 5 beat-to-beat intervals.
Heart Rate Alarms
: User-selectable.
Size
: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging.
Heart Rate Meter Response Time
:
Responds to a 40 BPM step increase in heart rate within 4.5 seconds per
AAMI EC-13-2002, section 4.1.2.1.f. Responds to a 40 BPM step decrease within
3.9 seconds per AAMI EC-13-2002, section 4.1.2.1.f. Response times include a
1.0-second display update interval.
Содержание Propaq MD
Страница 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Страница 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Страница 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Страница 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Страница 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Страница 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Страница 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Страница 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Страница 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Страница 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Страница 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Страница 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Страница 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Страница 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Страница 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Страница 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Страница 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...