Specifications
A-16
www.zoll.com
9650-0806-01 Rev. D
Alarms
Heart Rate Alarms:
Audible
: 5 pulse, 900 Hz tone, with a PW of 125 msec, a PRI of 250 msec, and a repetition
interval of 15 seconds.
Visual
: Heart Rate Alarm causes the heart rate to be displayed in red, with a white background.
The red device status LED will flash a rate of 1.7 Hz.
Lead Fault Alarm:
Audible
: 3 pulse, 500 Hz, triplet tone with a PW of 200 msec, a PRI of 200 msec. The lead
fault tone repeats at a repetition interval of 30 seconds.
Visual
: Lead Fault condition causes a LEAD FAULT message to be displayed on the trace
along with a dashed line the width of the trace.
Physiological Alarms (NIBP, SpO
2
, Resp, CO
2
, IP & Temp):
Audible
: Same as Heart Rate Alarm
Visual
: Physiological alarms cause the alarming parameter to be displayed in Red with a white
background. The red device status LED will flash at a rate of 1.7 Hz.
Silence Duration
: 90 seconds.
Invalid Operation Alert Tone:
A short, low-pitched tone is audible when a selected control button is unavailable for use or an
invalid entry is detected. Tone frequency is 160 Hz. Duration is 250 msec.
Delays (hold off time):
Heart rate/pulse rate
:
•
i
f source is ECG, 5 seconds
•
if source is SpO
2
, 10 seconds
•
if source is IBP, 5 seconds
•
if source is NIBP, no hold off
SpO
2
, SpCO, and SpMet Saturation:
10 seconds
EtCO
2
:
7 seconds
FiCO
2
:
5 seconds
IBP (Systolic, Diastolic, Mean):
2 seconds
Temperature:
2 seconds
Содержание Propaq MD
Страница 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Страница 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Страница 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Страница 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Страница 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Страница 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Страница 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Страница 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Страница 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Страница 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Страница 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Страница 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Страница 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Страница 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Страница 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Страница 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Страница 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...