Recorder
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
A-17
Recorder
Type
: High-resolution thermal array.
Annotation
: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters
and treatment summary events.
Paper Width
: 80 mm.
Paper Speed
: 25 mm/sec, 50 mm/sec
Delay
: 6 seconds.
Frequency Response
: Automatically set to monitor's frequency response.
Treatment Summary
:
10 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.).
Automatically logs into memory the type of event, time and ECG sample.
Full Disclosure Case Log
: A combination of 32 monitor snapshots; 500 non-ECG events; and
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance. The actual information stored can be more or less depending on the use profile and
the log configuration settings.
Record Modes
: Manual and automatic (User-configurable).
Battery
Type
: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh
Capacity
:
With a new, fully charged battery operating at room temperature:
•
At least
6 hours of continuous monitoring of ECG, SpO
2
, CO
2
, three Invasive Pressure
channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes and
10 200 J shocks (display set to 30%).
•
At least 100 discharges at maximum shock energy (200 joules).
•
At least 3.5 hours pacing, with ECG, SPO
2
, CO
2
, three Invasive Pressures, temperature,
NIBP every 15 minutes and pacing at 180 ppm, and 140 mA.
•
At least 10 discharges at maximum shock setting (200 joules) after a Low Battery
indication.
Note:
Proper battery care is required to maintain maximum available capacity.
Battery Indicators
:
5 Battery capacity LED indicators, Fault indicator, Recalibration indicator
Recharge Rate
: 100% in 4 hours, when initiated at Low Battery indication.
Содержание Propaq MD
Страница 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Страница 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Страница 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Страница 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Страница 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Страница 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Страница 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Страница 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Страница 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Страница 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Страница 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Страница 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Страница 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Страница 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Страница 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Страница 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Страница 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...