C
HAPTER
1 G
ENERAL
I
NFORMATION
1-10
www.zoll.com
9650-0806-01 Rev. D
motion and motion conditions, and for patients who are well or poorly perfused, in hospitals,
hospital-type facilities, or in mobile environments.
Respiration Monitoring
The Propaq MD is intended for use to continuously monitor respiration rate and to alarm if the
rate falls outside of the range set by the operator. Because the measurement method actually
measures respiratory effort, apnea episodes with continued respiratory effort (such as
obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The
patient population will range from newborn (neonate) to adult.
CO
2
Monitoring
The Propaq MD is intended for use to make continuous noninvasive measurement and
monitoring of carbon dioxide concentration of the expired and inspired breath and breath rate.
The patient population will range from newborn (neonate) to adult.
Invasive Pressure Monitoring
The Propaq MD is intended for use to display and make continuous invasive pressure
measurements from any compatible pressure transducer. The primary intended uses are arterial
blood pressure, central venous pressure and intracranial pressure monitoring. Any contra-
indications of the particular transducer selected by the user shall apply. The patient population
will range from newborn (neonate) to adult.
12-Lead Analysis
The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data,
and to provide interpretation of the data for consideration by caregivers. The interpretations of
ECG data offered by the device are only significant when used in conjunction with caregiver
overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis
is intended for use on adults (> 18 years of age).
Propaq MD Product Functions
Defibrillator Function
The Propaq MD contains a direct current (dc) defibrillator capable of delivering up to 200
joules. It may be used in synchronized mode to perform synchronized cardioversion using the
patient’s R-wave as a timing reference. The unit uses paddles or disposable, pregelled
electrodes for defibrillation.
Содержание Propaq MD
Страница 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Страница 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Страница 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Страница 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Страница 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Страница 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Страница 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Страница 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Страница 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Страница 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Страница 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Страница 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Страница 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Страница 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Страница 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Страница 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Страница 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...