Defibrillator/Pacing Test with Hands-Free Therapy Electrodes
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
24-3
Defibrillator/Pacing Test with Hands-Free Therapy
Electrodes
Note:
If a low battery message appears during any of this testing, the battery is close to
depletion and should be replaced or recharged.
Warning!
Keep hands, fingers, and other conductive materials away from paddle electrode plates
when discharging the defibrillator, or pacing.
Function
Response
1
Connect the external AC adapter to a working
AC outlet and to the Propaq MD rear panel.
Verify that the green Auxiliary Power LED
illuminates on the Propaq MD front panel.
2
Insert a battery into the unit (if a battery is not
already in the slot).
Verify that the Battery Charge LED on the
Propaq MD front panel illuminates yellow.
When the battery pack has been fully charged,
the battery status indicator light is green.
3
Press the unit's power on switch.
Verify the unit issues two audio beeps and that
the red, yellow, and green Visual Alarm
indicators illuminate for 2 to 3 seconds.
Note: The yellow LED may remain lit after the
unit powers on.
The Propaq MD unit performs a number of
power on self tests.
Verify that the unit displays SELF TEST
PASSED.
If the message SELF TEST FAILED appears
on the display, contact the appropriate
technical personnel or the ZOLL Technical
Service Department.
4
Disconnect the external AC adapter from the
Propaq MD rear panel.
Verify that the unit continues to operate without
interruption on battery power and that the
displayed battery icon indicates that a fully
charged battery is installed.
Verify that no error messages are displayed.
5
With no ECG cables attached to the unit,
select Lead I, II, or III as the waveform display
source.
Verify that a LEAD FAULT message displays in
the waveform window.
6
Select Pads as the waveform display source
and connect the multifunction cable to the
unit. (Do not connect the test connector to the
MFC.)
Verify that a CHECK THERAPY
ELECTRODES message displays in the
waveform window.
7
Connect the multi-function cable (MFC) to the
test connector.
Verify that the CHECK THERAPY
ELECTRODES message is replaced by a
SHORT DETECTED message.
Содержание Propaq MD
Страница 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Страница 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Страница 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Страница 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Страница 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Страница 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Страница 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Страница 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Страница 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Страница 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Страница 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Страница 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Страница 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Страница 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Страница 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Страница 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Страница 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...