Page
2-9
User Guide
2.2
Battery Installation and Removal
Remove a charged Battery from the Battery Charger. Ensure the Battery is fully charged before insertion
into the Platform (refer to the
AutoPulse Power System User Guide
for more information). Replace the
Battery remaining in the Platform with one that is fully charged.
2.2.1
AutoPulse Li-Ion Battery Installation and Removal
Figure 2-11
AutoPulse Li-Ion Battery Installation and Removal
To install the Li-Ion Battery, first make sure the Battery Latch is rotated out of the way so that the Li-Ion
Battery can slide into the Battery compartment (see Figure 2-11). Then slide the Li-Ion Battery into the
Battery compartment in the Platform. The Li-Ion Battery should snap into place and mount flush with
the Platform. The Finger Latch for the Li-Ion Battery should also be flush with the Platform such that
the red strip on the inside of the Finger Latch is not visible. Then rotate the Battery Latch into position
as needed.
The Li-Ion Battery is mechanically keyed so that it can only be inserted in one orientation. If resistance
is met, check for appropriate orientation, and check to ensure there are no obstructions to battery
insertion and that the Battery Latch is open.
Finger Latch
Battery Latch Opened
Battery Latch Closed
Battery Latch
Locking Bar
Battery Front
Содержание AutoPulse 100
Страница 1: ...P N 12555 001 Rev 9 Resuscitation System Model 100 User Guide...
Страница 8: ...Page viii User Guide This page left intentionally blank...
Страница 38: ...Page 2 16 User Guide This page left intentionally blank...
Страница 66: ...Page 4 6 User Guide This page left intentionally blank...
Страница 72: ...Page 5 6 User Guide This page left intentionally blank...
Страница 74: ...Page A 2 User Guide This page left intentionally blank...