Page 2-2
User Guide
5.
Make sure that the band clip is seated properly and fully into the slot on the driveshaft (see Figure
2-2).
Note:
If the band clip is properly seated, you should be able to turn the driveshaft in each
direction by hand.
Figure 2-2
Seating the LifeBand Band Clip Properly into the Driveshaft Slot
6.
Ensure that both free ends of the LifeBand are oriented flat (not twisted) and away from the
AutoPulse Platform. Inspect the LifeBand for any cuts or tears. Do not use the LifeBand if cuts or
tears are present.
7.
Line up the arrow on the LifeBand cover plate with the matching arrow on the Platform.
8.
Snap the LifeBand cover plate in place by fully inserting the locking tabs into the slots on the
Platform (see Figure 2-3).