Errors
13
Zolix Instruments Co., Ltd.
http://www.zolix.com.cn E-mail:[email protected]
6. Errors
Phenomenon
Possible Reasons
Ways and means
The stage doesn’t move
1. Power off
2. Cable connection error
3. The stage has not been
moved to the zero position.
4. Direction error
1. Turn on the power of the
SC300
2. Connect the cable again
3. Firstly move the stage to
the zero position
4. Select the movement
direction again
The displacement is not
the setting
Input error steps
Input correct steps again
The speed of the stage is
not correct
1. Select the error speed.
2. There is something in the
screw or guide.
3. Low lubrication
4. Overloading
1. Input the correct speed
again
2. Return the product back to
Zolix
3. Add lubrication
4. Decrease the load
Big noise
1. There is something in the
screw or guide.
2. Low lubrication
3. The stage is not fixed up
tightly on the table.
1. Return the product back to
Zolix
2. Add lubrication
3. Tighten screws
The stage cannot stop
after touching the limit
switch
The limit switch is invalid
1.Check up the line on the
limit switch
2. Return it back to Zolix.
After sending ”HX
↓
”, the
stage move in reverse
direction and doesn’t stop
while the limit switch is
reached.
The limit switch on the zero
position is faulty.
1.Check up the line on the
switch
2. Return it back to Zolix.
The stage moving in one
direction is not correct.
1. The driver is faulty.
2. The controlling software is
faulty.
1. Change the driver.
2. Amend the software
3. Use the Zolix’s software