Measuring system XS22-E2010
9 Use of ZrO
2
-probes
21
XS22_E2010
9 Details for the use of ZrO
2
-probes in vacuum technology
9.1 Theoretical relations
Potentiometric ZrO
2
-probes measure the difference of the oxygen partial
pressure at both electrodes.
One electrode is fixed inside, the other one outside the ZrO
2
-tube. Air as
reference gas flows around the inner electrode.
The NERNST equation applies:
)
(
)
(
ln
4
2
2
gas
measuring
p
air
p
F
RT
U
O
O
=
(Nernst equation)
Condition: isothermal electrodes
U
– cell voltage
R
– molar gas constant, R = 8.31 J/(mol
.
K)
T
– Kelvin temperature
F
– Faraday constant, F = 9.64
.
10
4
C/mol
p
O2
(air) – partial pressure of oxygen at reference electrode
p
O2
(measuring gas) – partial pressure of oxygen at measuring electrode
After inserting the known values of the natural constants the following applies:
]
/
/
42
.
46
exp[
)
(
)
(
2
2
K
T
mV
U
gas
reference
p
gas
measuring
p
O
O
⋅
−
⋅
=
]
/
/
42
.
46
exp[
2064
.
0
)
(
2
K
T
mV
U
p
gas
measuring
p
L
O
⋅
−
⋅
⋅
=
0.2064 refers to the oxygen concentration in air at 50 % relative humidity and
p
L
is the air pressure. The pressure unit of p
O2
(measuring gas) and p
L
can
be chosen user-defined but must be identical for both pressures. The
numerical value in this equation requires the probe signals U used in mV and
T in Kelvin.
For the calculation of the oxygen concentration
ϕ
O2
in vacuum, which is the
ratio of the number of oxygen molecules to the total of gas particles, the total
pressure in the vacuum chamber is necessary.
Содержание XS22
Страница 2: ...2 XS22_E2010...
Страница 4: ...4 XS22_E2010...
Страница 20: ...20 XS22_E2010...
Страница 24: ...24 XS22_E2010...
Страница 26: ...26 XS22_E2010...
Страница 27: ...Measuring system XS22 E2010 11 Declaration of conformity 27 XS22_E2010 11 Declaration of conformity...