Measuring system XS22-E2010
5 Installation
13
XS22_E2010
NOTE
5 Installation and initiation
5.1 Installation
Connect the circuit points on the rear of the device with the corresponding
points of the ZIROX-sensor or the ZIROX-probe.
The operating voltage is 24 V DC.
5.2 Reference air supply for ZIROX
®
probes
This paragraph only applies to special probe constructions, which require
reference air supply because of their length.
Reference air input of the sensor and the reference air output of the E2010
must be connected by a flexible hose (inner diameter 3 mm). The reference
air flow is preset by the manufacturer.
After starting the system (E2010 with connected sensor) the reference air
flow must be checked by an inserted flow meter (5 - 10 l/h resp. approx. 100
- 200ml/min) and, if applicable, adjusted via display.
5.3 Probe mounting
The probes are mounted with a flange DN40 KF (according to order, others
on request).
When pulled out after a running process the probe is very hot. Burn and fire
hazards!
5.4 Waiting for operating state
Depending on the connected sensor the E2010 needs a certain period of
time until the operating state is reached (for further information see technical
data of sensor). Due to thermal balancing effects, the sensor needs approx.
60 minutes until the ultimate operating state after reaching the operating
temperature.
The keys are locked before reaching the operating state!