9 Use of ZrO
2
-probes
23
XS22_E2010
In practice, if the probe is needed for the adjustment and replication of
defined conditions only, it is sufficient to determine the probe signals
empirically in a successful trial. The operating mode has to be set up so as
to receive the requested probe signal. Due to different parameter impacts on
the raw signal of the probe this is the easiest and in some case the only way.
Receiving theoretical values is only possible under highly simplified (lab)
conditions for the most part.