Manufacturer's Declaration on
Electromagnetic Compatibility (EMC)
15
Page 27
Table 1
Guidance and Manufacturing Declaration- Electromagnetic Emissions
The device MFG-03 is intended for use in the electromagnetic environment specified below. The customer or user of
the device MFG-03 should ensure that it is used in such environment.
Emission Measurement
Compliance
Electromagnetic Environment-Guidelines
RF Emissions in accordance
with CISPR 11
Group 1
The device MFG-03 must emit electromagnetic
energy in order to ensure its intended function.
Nearby electronic equipment may be affected.
RF Emissions in accordance with
CISPR 11
Class A
The quality of the supply should correspond to that of
a typical business or hospital environment. If the user
of the MFG-03 requires continued function even if
interruptions in the power supply occur, it is
recommended to power MFG-03
from an
uninterruptible power supply or a battery.
Emissions of Harmonics in
accordance with IEC 61000-3-2
Class A
Emissions of voltage fluctuations/
flickers in accordance with IEC
61000-3-3
Not performed
Table 2
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device MFG-03 is intended for use in the electromagnetic environment specified below. The customer or user of
the device MFG-03 should ensure that it is used in such environment.
Immunity
Tests
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment -
Guidelines
Electrostatic
Discharge
(ESD) in
accordance
with IEC
61000-4-2
± 8 kV Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV Air Discharge
± 8 kV Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV Air Discharge
Floors should be made from wood,
concrete or ceramic tiles. If floors are
covered with synthetic material, the
relative humidity must be at least 30 %
Electrical
fast
transient/
burst in
accordance
with IEC
61000-4-4
± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition frequency
The supply voltage quality must
correspond to that of a typical
commercial or hospital
environment.
Surges in
accordance
with IEC
61000-4-5
-Line-to-
Line-
± 0,5 kV, ± 1 kV
± 0,5 kV, ± 1 kV
Surges in
accordance
with IEC
61000-4-5
-Line-to-
Earth-
± 0,5 kV, ± 1 kV, ± 2 kV
± 0,5 kV, ± 1 kV, ± 2 kV
Содержание Z Field Dual
Страница 1: ...Instructions for use ZFieldDual MFG 03 EN...
Страница 2: ......
Страница 6: ...Illustrations Accessories 21 Mains cable 22 Fixingbelt...
Страница 40: ...Material Safety Data Sheet 16 Page 30...
Страница 41: ...Material Safety Data Sheet 16 Page 31...
Страница 42: ...Material Safety Data Sheet 16 Page 32...
Страница 43: ...Material Safety Data Sheet 16 Page 33...
Страница 44: ...Material Safety Data Sheet 16 Page 34...
Страница 45: ...Material Safety Data Sheet 16 Page 35...
Страница 46: ...Material Safety Data Sheet 16 Page 36...