Manufacturer's Declaration on
Electromagnetic Compatibility (EMC)
15
Page 28
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device MFG-03 is intended for use in the electromagnetic environment specified below. The customer or user of
the device MFG-03 should ensure that it is used in such environment.
Immunity
Tests
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment -
Guidelines
Voltage dips
in
accordance
with IEC
61000-4-11
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
The supply voltage quality must
correspond to that of a typical
commercial or hospital environment. If
the user of the device MFG-03
requires
continued operation, even in the case of
interruptions in the power supply, it is
recommended that the device MFG-03
be powered from an uninterrupted
power supply or a battery.
0 % U
T;
1 cycle and 70% U
T;
25/30 cycles
Single phase: at 0°
0 % U
T;
1 cycle and 70% U
T;
25
cycles
Single phase: at 0°
Voltage
interruptions
accordance
with IEC
61000-4-11
0% U
T;
250/300 cycles
0% U
T;
250 cycles
Magnetic
field of
supply
frequency
(50/60 Hz) in
accordance
with IEC
61000-4-8
30 A/m
50 Hz oder 60 Hz
30 A/m
50 Hz
Magnetic fields at mains frequency
should have the typical values found in
a business or hospital environment.
Note: U
T
is the mains AC Voltage before application of the test level
Table 3
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device MFG-03 is intended for use in the electromagnetic environment specified below. The customer or user of
the device MFG-03 should ensure that it is used in such environment.
Immunity Test
IEC 60601-Test
Level
Compliance Level
Electromagnetic Environment -
Guideline
Conducted
Disturbances induced
by RF fields
according IEC 61000-
4-6
3 V
0,15 MHz to 80 MHz
6 V in ISM Band
between 0,15 MHz
and 80 MHz
80% AM at 1 kHz
3 V
0,15 MHz to 80 MHz
6 V in ISM Band between 0,15
MHz and 80 MHz
80% AM at 1 kHz
In the vicinity of devices, bearing the
following symbol, interference is
possible:
Radiated RF EM fields
according IEC 61000-4-
3
3 V/m
80 MHz-2,7 GHz
80% AM to 1 kHz
3 V/m
80 MHz-2,7 GHz
80% AM to 1 kHz
Содержание Z Field Dual
Страница 1: ...Instructions for use ZFieldDual MFG 03 EN...
Страница 2: ......
Страница 6: ...Illustrations Accessories 21 Mains cable 22 Fixingbelt...
Страница 40: ...Material Safety Data Sheet 16 Page 30...
Страница 41: ...Material Safety Data Sheet 16 Page 31...
Страница 42: ...Material Safety Data Sheet 16 Page 32...
Страница 43: ...Material Safety Data Sheet 16 Page 33...
Страница 44: ...Material Safety Data Sheet 16 Page 34...
Страница 45: ...Material Safety Data Sheet 16 Page 35...
Страница 46: ...Material Safety Data Sheet 16 Page 36...