Cleaning
Disinfection
8
Page 17
Monitoring:
1.
Check whether the fixingbelts are free from punctures or other damaged areas.
2.
Check whether the fixingbelts show uniform thickness over the entire surface.
Packaging:
Pack the fixingbelts in cloth or foil pouches or other suitable dust cover to protect them from
dust and dirt.
Reusability:
Over time, it can – depending on use and treatment – come to the destruction of the material.
Once the fixingbelts are punctured, worn, have optical damages, have non-removable
contaminations or the Velcro
TM
is not functional anymore, please do not reuse them.
In case the label is torn off, please do not reuse the fixingbelt.
Housing/applicator
Cleaning: if there is visible contamination, the housing, the applicator and all cables can be
cleaned with commercially available, mild, non-alcoholic plastic cleaners. Wipe the surface
with a soft cloth that has been soaked in the detergent according to the manufacturer's
instructions, but is not dripping wet, until the dirt is removed.
Disinfection: we recommend disinfecting at a minimum after each patient and at the end of
the day and whenever there is evidence of possible contamination. Contact your health
expert on this topic. Always perform cleaning prior to disinfection.
The housing and applicator can be disinfected with disinfectant wipes. For metal and plastic,
use a commercially available, non-alcoholic disinfectant with bactericidal, virucidal and
fungicidal properties. Please observe manufacturer's instructions for use. Wipe all surfaces
with a cloth soaked in the disinfectant according to the manufacturer's instructions, but not
dripping wet, or with a cloth that is pre-impregnated with disinfectant.
If applicable, also observe the requirements for drying or subsequent cleaning.
Caution: If inflammable solutions are used for cleaning and disinfection, sufficient time must
be allowed for the solutions to evaporate before using the device. Otherwise fires may be
caused!
Note:
Only use the device in a hygienic environment.
Содержание Z Field Dual
Страница 1: ...Instructions for use ZFieldDual MFG 03 EN...
Страница 2: ......
Страница 6: ...Illustrations Accessories 21 Mains cable 22 Fixingbelt...
Страница 40: ...Material Safety Data Sheet 16 Page 30...
Страница 41: ...Material Safety Data Sheet 16 Page 31...
Страница 42: ...Material Safety Data Sheet 16 Page 32...
Страница 43: ...Material Safety Data Sheet 16 Page 33...
Страница 44: ...Material Safety Data Sheet 16 Page 34...
Страница 45: ...Material Safety Data Sheet 16 Page 35...
Страница 46: ...Material Safety Data Sheet 16 Page 36...