background image

CA103&CA102S Handy Multi-functions Calibrator Operating Instruction

 

12 

 

• When  using  the  instrument  in  ambient  humidity  of  30%  or  less,  use  an 

antistatic mat to prevent static electricity. 

• Condensation may occur if the instrument is moved from a location of low 

temperature and humidity to a location of high temperature and humidity, or 

if  the  temperature  otherwise  changes  suddenly.  In  such  a  case,  leave  the 

instrument  for  at  least  one  hour  to  ensure  it  is  free  from  condensation 

before starting operation 

 
3.2 Connecting a Power Supply 

Calibrator can use two other types of power supply. 

• AC adapter 
• NiMH (nickel hydrogen) battery(6 battery) 

3.2.1Using an AC Adapter 

Ratings 

 

AC adapter power supply ratings 

(Model

CA-P01) 

Rated supply voltage 

100~240V AC 

Rated supply frequency 

50/60Hz 

Maximum input current 

0.5A 

Output voltage rating of AC adapter 

12V DC 

Maximum output current rating of AC adapter 

1.0A 

3.2.2 Using an NiMH Battery Pack 

Charging type   NiMH (nickel hydrogen) battery (optional): Model: CA-C02 

Specifications     Voltage: 1.2V

×

6 V 

Capacity             1600 mAh  

Number of times can be charged (life cycle): Approx. 300 

times  (varies  depending  on  the  operating  environment) 

Be  sure  to  observe  the  following  warnings  on  handling 

the NiMH battery. 

 

http://www.zhoyi.com

Содержание CA103

Страница 1: ...CA103 CA102S Handy Multi Functions Calibrator User s Manual http www zhoyi com Shaoxing Zhongyi Electronics Co Ltd...

Страница 2: ...s strictly prohibited The warranty is included in this manual Be sure to read the warranty to ensure you understand the terms and then store it in a safe place The warranty cannot be reissued Revision...

Страница 3: ...the instrument Main Unit Check that the model name given on the name plate on the back panel of the instrument matches the one on your order Model Model CA103 CA102S Serial No Should you need to cont...

Страница 4: ...that may result in the loss of life or serious injury of the user unless the described instruction is abided by Caution Indicates a hazard that may result in an injury to the user and or physical dam...

Страница 5: ...any external circuit turn off the power to the circuit and make sure that no voltage is generated Correct Use of Lead Cables Be sure to correctly use the measurement lead cables and source lead cable...

Страница 6: ...iMH Battery Pack 12 3 3 TURNING THE POWER ON AND OFF 14 3 3 1 Turning the Power On and Off 14 3 3 2 Screen saver 14 3 3 3 Auto Power Off 14 3 4 OPERATING ENVIRONMENT 14 4 SOURCE 15 4 1 CONNECTING THE...

Страница 7: ...RE DETECTOR RTD 23 5 5 MEASURING THERMOCOUPLE TC 24 5 6 MEASURING FREQUENCY FREQ 25 6 PARAMETERS SETTING 25 6 1 GENERAL PARAMETERS SETTING 26 6 2 SECOND LEVEL CALIBRATION PARAMETER SETTING 27 7 NTERNA...

Страница 8: ...DCV 100mV 6V DC Current DCA 24mA 24mA SINK Resistance 500 Thermocouple TC S K E T J B R N WRe3 25 Frequency FREQ 1000Hz 5kHz Measurement MEASURE Function Range DC Voltage DCV 30V 100mV DC Current DCA...

Страница 9: ...on 8 2 Names and Functions of Parts Output Range Output Function Output Value Additional Information Power Key Numeric Key Pen Outlet Power Indicator Charge Indicator Function key Direction key Measur...

Страница 10: ...h to the follow the sequence S K E T J B R N W325 Measure Keys Changes the measure function The measurement sequence is DCV DCmA DCmV RTD TC F DCV Changes the range for each function Common Keys Outpu...

Страница 11: ...revent heat sources from coming into contact with them When unplugging the power cord from the outlet be sure to hold the plug and never pull the actual cord If the power cord is damaged contact your...

Страница 12: ...locations In direct sunlight or near heat sources Where the instrument is exposed to water or other liquids Where there is a lot of mechanical vibration Near noise sources such as high voltage equipme...

Страница 13: ...Power Supply Calibrator can use two other types of power supply AC adapter NiMH nickel hydrogen battery 6 battery 3 2 1Using an AC Adapter Ratings AC adapter power supply ratings Model CA P01 Rated su...

Страница 14: ...upture Do not short the NiMH battery as this may cause burns due to the battery pack heating up Do not dispose of the battery pack in a fire or apply heat to it otherwise there is a risk that it will...

Страница 15: ...et time if no operation the LCD into the screen saver press any key to exit the screen saver status LCD Screen Saver parameter value is set to 0 to disable this function 3 3 3 Auto Power Off After set...

Страница 16: ...tion 4 1 Connecting the Source Terminals Connect the supplied source lead cables model CA B01 to the output terminals of the instrument Connect the clips Alligator clip CA B02 to the input terminals o...

Страница 17: ...nt output value 1 5V display When data is beyond the range will not display When the output error exceeds output error parameter settings display error 3 Use the Direction keys or numeric keys to set...

Страница 18: ...ernal voltage source distributor etc in the direction of the H terminal Thus you can use the instrument in a loop test for example as a simulator for two wire transmitters Note When using 20mA Sink fu...

Страница 19: ...n the output interface is marked as DCmV DC current choice 3 Use the Ra Keys to select the type and range SINK current SINK 1 24mA In the additional information column corresponds to the percentage of...

Страница 20: ...ys to set the output value Press the Zero key return to the initial value is set to 10 4 7 Source Thermocouple TC Signals Generate voltages mV corresponding to the following thermocouples Set the temp...

Страница 21: ...ue is set to 0 5 Measurement Warning Turn off the power supply to the object to be tested before connecting it to the instrument It is extremely dangerous to connect and disconnect measurement lead ca...

Страница 22: ...g DC Voltage DCV 1 Connect the terminals as shown in the figure above 2 Use the Measure key on the Fun when the output interface is marked as DCV DC voltage measurement range is displayed CA103 Ran 30...

Страница 23: ...and unit mA appear 5 2 2 Measuring 24V LOOP This function is for measuring the current flowing when a 24 V DC constant voltage is applied With 24V LOOP measurement you can perform a transmitter loop t...

Страница 24: ...minals as shown in the figure above 2 Use the Measure key on the Fun when the output interface is marked as RTD RTD choice 3 Use the Ra Key to select the type of thermal resistance Three WireCu50 Thre...

Страница 25: ...e range refer to the specifications 1 Connect the terminals Figure 4 1 2 Use the FUNCTION key on the SOURCE when the output interface is marked as TC Thermocouple selection 3 Use the RANGE key to sele...

Страница 26: ...he figure above 2 Use the Measure key on the Fun when the output interface is marked as FREQ FREQ measurement range is displayed Ran 0 5KHz 3 The measurement value and unit Hz appear 6 Parameters sett...

Страница 27: ...ter modification screen shown below 4 Press the value of the ENT key to save the input parameters and return items to see the picture If you press ESC key to return without saving 5 After parameters s...

Страница 28: ...9999 19999 00000 DC mA offset for cal DCmA Prop 0 00001 9999 1 0000 DC mA proportion for cal Compensation Temperature 00 0 99 9 00 0 Cold junctions compensation value of T C input measuring in manual...

Страница 29: ...ortion for cal TC JZero 19999 19999 00000 J type TC offset for cal TC JProp 0 0000 1 9999 1 0000 J type TC proportion for cal TC BZero 19999 19999 00000 B type TC offset for cal TC BProp 0 0000 1 9999...

Страница 30: ...set to zero in the corresponding parameters 2 Output error parameter settings 0 Disable output error alarm function 1 200 Enable output error alarm function When the signal source output error output...

Страница 31: ...Green LED lights up while the Ni MH battery pack is charged Red LED lights up while the Ni MH battery pack is Fast charged indicate Red LED Green LED Charging time Minute Charging Light on Flicker 0 5...

Страница 32: ...1372 0 0 03 0 5 0 1 TC source accuracy does not include RJ sensor accuracy RJ Sensor Specification Measurement range 10 50 Accuracy when combined with main unit 18 28 0 5 Other than 18 28 1 E 0 0 100...

Страница 33: ...ment 5kHz 0 0 5000 0Hz 0 1 2dgt 0 1Hz Maximum input 30V Sensitivity Approx 0 5Vpp TC input K E J T N We3 25 50 0 1372 0 50 0 1000 0 50 0 1200 0 50 0 400 0 50 0 1300 0 50 0 2300 0 0 03 1 5 100 0 0 03 2...

Страница 34: ...0 1 TC source accuracy does not include RJ sensor accuracy RJ Sensor Specification Measurement range 10 50 Accuracy when combined with main unit 18 28 0 5 Other than 18 28 1 E 0 0 1000 0 0 05 0 4 J 0...

Страница 35: ...x 0 5Vpp TC input K E J T N We3 25 50 0 1372 0 50 0 1000 0 50 0 1200 0 50 0 400 0 50 0 1300 0 50 0 2300 0 0 05 1 5 100 0 0 05 2 100 0 0 1 R S B 50 0 1768 0 50 0 1768 0 50 0 1820 0 0 05 3 RTD input Cu5...

Отзывы: