CA103&CA102S Handy Multi-functions Calibrator Operating Instruction
11
panel is made of thermoplastic resin, be careful not to let it come into
contact with any heat sources such as a soldering iron.
●
If the AC adapter will not be used for a prolonged period, unplug the
power cord from the outlet.
●
Do not use the instrument with the cover for the battery holder left open.
Operating Environment
Use the instrument in locations that meet the following conditions:
• Ambient temperature and humidity
Ambient temperature: 0 to 40
℃
Ambient humidity: 20 to 80% RH (no condensation)
Do not use the instrument in the following locations:
• In direct sunlight or near heat sources
• Where the instrument is exposed to water or other liquids
• Where there is a lot of mechanical vibration
• Near noise sources such as high-voltage equipment or power lines
• Near strong magnetic field sources
• Where an excessive amount of greasy fumes, steam, dust, or corrosive
gases are present
• In an unstable place
• Where, for example, fire and explosions caused by inflammable gases
and the like are possible
☞
Note
• When you require high source and measurement accuracy, use the
instrument under the following conditions:
Ambient temperature: 23 ± 5
℃
Ambient humidity: 20 to 80% RH (no condensation)
When using the instrument in an ambient temperature range of 0 to
18
℃
or 28 to 40
℃
, add the temperature coefficient specified in "9.
Specifications" to the accuracy.
http://www.zhoyi.com