CA103&CA102S Handy Multi-functions Calibrator Operating Instruction
19
2 Use the FUNCTION key on the SOURCE ,when the output interface
is marked as DCmV, this time to display DC voltage signal output range
(Ran 0-100mV), in the additional information column when the output error
exceeds "output error" parameter settings, display "error".
3 Use the Direction keys or numeric keys to set the output value.
Press the Zero key, return to the initial value is set to "0."
4.6 Source Resistance (RESI) Signals
1 Connect the terminals Figure 4.1.
2 Use the FUNCTION key on the SOURCE ,when the output interface
is marked as RESI, this time to display DC voltage signal output range
(Ran 10-510
Ω ).
3 Use the Direction keys or numeric keys to set the output value.
Press the Zero key, return to the initial value is set to "10."
4.7 Source Thermocouple (TC) Signals
Generate voltages (mV) corresponding to the following thermocouples.
Set the temperature with (
℃
). This enables calibration of a thermometer.
Thermocouple (TC) types: S
,
K
,
E
,
T
,
J
,
B
,
R
,
N
,
W325 (For
the temperature range, refer to the specifications.)
1 Connect the terminals Figure 4.1.
2 Use the FUNCTION key on the SOURCE, when the output interface
is marked as TC. (Thermocouple selection)
3 Use the RANGE key to select a thermocouple type.
In the additional information column shows the current measured
temperature corresponds International Temperature Scale ITS-90 value in
millivolts. When the output error exceeds "output error" parameter settings,
display "error".
4 Use the Direction keys or numeric keys to set the output value.
Press the Zero key, return to the initial value is set to "0."
http://www.zhoyi.com