CA103&CA102S Handy Multi-functions Calibrator Operating Instruction
27
6.2 Second level (Calibration) parameter setting
Enter the correct password in the password entry proofreading generic
parameter settings, press ENT, then move to second level parameter
setting
☞
Note
•Non-engineering personnel shall not enter two parameters. Otherwise, it
will cause the instrument to measure the error!
Parameter
Range
Default
setting
Description
Password
0000
~
9999
0000
Password change or verify
DC-VZero
-19999+19999
00000
DC V Offset for cal
DC-VProp
0.00001.9999
1.0000
DC V proportion for cal
DCmAZero
-19999+19999
00000
DC mA offset for cal
DCmA Prop
0.00001.9999
1.0000
DC mA proportion for cal
Compensation
Temperature
00.0
℃
~99.9
℃
00.0
℃
Cold-junctions compensation value of
T.C input measuring in manual mode.
Output
Compensation
Manual or
Auto
Manual
Cold-junctions compensation value
for T.C output manual setting.
Temperature
Compensation
00.0
℃
~99.9
℃
00.0
℃
T.C input measuring cold-junctions
compensation for manual setting.
Contrast
0~10
05
Screen contract setting
screen saver
0
~
120
03
Screen saver timer time
(Unit: minutes)
Auto Power
Off
0
~
120
05
Auto Power Off timer time
(Unit: minutes)
Key Sound
Turn on or off
turn on
Sound switch button
Language
Chinese or
English
Chinese
English display settings
password
0000~9999
0000
For into the second level parameter
setting page