This operating manual describes the function and handling of the Radio Frequency
Surgery Equipment ZEUS-200S/150. The manual serves as an instruction reference
and should be read thoroughly before operating the equipment. Only then can the
correct handling of the equipment be assured. In case of incorrect handling no liability
will be taken on by the manufacturer.
2.1 Before Application
●
About Accessories
①
Electric Shock
Don't connect wet accessories to the surgical unit; precisely connect all accessories
to adapters; don't expose metal materials to the air.
②
Check the linkage condition between all accessories and Electrosurgical Unit before
using them. If the connection is bad, it can cause unintentional operation effects like
spark or accessory malfunction.
③
Don't wrap up the accessory code or Patient Return Plate in metal materials.
This can generate shock and fire, inflicting an injury on patients or operators.
④
Before using accessories, always check if reusable accessories and codes are not
broken, cracked, scratched, etc. If this is not done, patients or operators can be
injured or struck by electricity.
⑤
Connect accessories to the appropriate sockets because bad connection can cause
dangerous situations. Unused accessories (active electrode) must not be placed on
the patient body.
⑥
ZERONE Co., Ltd. Supplied accessories have appropriate for rating for ZEUS-200S/
150. CE marked or UL approved components with more than Monopolar: 1800Vpp,
Bipolar: 500Vpp rated voltage rating can be used as accessories.
If the rating of the accessory is lower than the max output of the applicable mode,
set up the output as much as the rating of the applicable accessory.
Safety hazard may occur if products with lower quality are utilized.
If any malfunction or breakage. do not use it for patients and instead contact
with medical equipment technicians or suppliers.
Warning
Check the followings before using this product.