ZEUS-200S/150 ZO-OM-007
REV : 08 11 ZERONE
⑨
Extreme caution must be taken for patients with pacemaker or implantable electrode.
Performance of devices may be temporarily affected which may cause ventricular
fibrillation. In case of any doubts concerning safety of patient contact qualified and
concerned pacemaker manufacturer or Cardiology department.
⑩
Do not use Active Electrode near to Electrocardiograph electrode. ESU electrode
must be kept apart at least 150mm from ECG electrode. It is necessary to use
Monitoring Electrode with protective resistors to survive high frequency. Do not use a
needle-typed electrode.
⑪
Most sweated parts of human body (armpit, knee, skin-to-skin parts) can be burned
when Patient Return Plate or other materials contacts with those parts, therefore,
some dried and absorbable towels should be used to dry them.
⑫
Lower output or performance of electrical surgical device despite correct surgical
setting may be caused due to incorrectly applied patient plate or bad connection.
In this case, before setting a higher output, you must check the connection status.
⑬
The output from electrodes may change during use.
⑭
A failure of the electrosurgical unit could result in an unintended increase of output
power.
①
The patient leads should be positioned in such a way that contact with the patient or
other leads is avoided. Temporarily unused electrodes should be stored in a location
that is isolated from the patient.
②
For surgical procedures where the HF current could flow through parts of the body
having a relatively small cross sectional area, the use of BIPOLAR techniques may
be desirable in order to avoid unwanted tissue damage.
●
About Patient Return Plate
①
Plate Electrode should be near to an operation part of human body and adhere closely
to the patient
’
s body. Periodically check if the electrode firmly contacts to the body
when moving the patient or performing a long operation.
②
Contact Plate Electrode closely to the place whose direction is the opposite from the
heart based on the operated part.
Check the followings before using this product.
Warning
Check the followings before using this product.
Caution