ZEUS-200S/150 ZO-OM-007
REV : 08 14 ZERONE
⑬
If the area between a patient and the pad is not appropriate, REM(Return Electrode
Monitoring) gives the alarm with a warning sound and stop an operation of the product
to prevent a burning incident.
⑭
The level of alarm sound can be adjusted during an operation of application mode.
⑮
The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed
by reenergizing the device after power OFF.
5.2
Safety Function
①
A fuse built in a power circuit prevent an over current from flowing through the
equipment.
②
When the plate attached to the patient is separated from the equipment, the red
alarm light begins flickering. Pressing the button of Twin Button Handle or the
pedal of Cutting and Coagulation of Double Foot Switch will stop the alarm sound
and the equipment.
③
REM (Return Electrode Monitoring) monitors the size of contacting area between a
patient and the pad. If the size is inappropriate, it automatically blocks the high-
frequency current to minimize the danger of burning incidents.
④
To protect a patient, the case is fully grounded so that a leakage current can flow
into the earth.
5.3
Indication for use
Electrosurgical unit is a device to perform a medical operation like cut and
coagulation in the biological tissue using high-frequency current
●
Operator Profile
Considerations
Requirement description
Education
Minimum
A physician or medical personnel under the
supervision of a physician
Maximum
N/A
Language
understanding
Minimum
Languages specified in the instruction for use
Maximum
N/A
Experience
Minimum
A person who received sufficient training in
electrosurgical procedures
Maximum
N/A
Permissible
impairments
N/A