ZEUS-200S/150 ZO-OM-007
REV : 08 29 ZERONE
9.
Troubles and What to do with them
●
There are no lights on the Display Number window.
1. Check if the power cord is connected to the power cord input plug on the back
side of the surgery unit.
2. Check if the power switch of the surgery unit is on.
3. Check the fuse (AC120V: T6.3AL 250V/ AC230V: T4.0AL 250V) on the back side
of the equipment.
4. If the trouble continues, use the auxiliary equipment.
●
No output.
1. Check if the plug of patient return plate cable is connected to the surgery unit.
2. Check if the patient return plate is accurately connected with the patient.
3. Check if the accessories (Twin Button Handle, Monopolar Handle and Foot Switch)
are connected.
4. Change the accessories such as Twin Button Handle, Monopolar Handle and Foot
Switch and others.
5. Check if the output is set low.
6. If the trouble continues, use the auxiliary equipment.
●
The alarm rings all the time.
1. When the patient plate has troubles.
①
Check if the plug of patient plate cable is connected to the surgery unit.
②
Change the patient plate.
③
If the trouble continues, use the auxiliary equipment.
2. When the patient return plate has troubles.
①
Check if the plug of patient return plate cable is connected to the surgery unit.
②
Check if the entire surface of patient return plate is adhered to the patient.
③
Check if the patient return plate is connected to the patient return plate cable.
④
Change the patient return plate cable.
⑤
If the trouble continues, use the auxiliary equipment.