ZEUS-100/80 ZO-OM-008
REV : 08 21
ZERONE
5.6
Using Method and Procedure
●
Monopolar
①
After confirming power and voltage (AC120V/AC230V), connect a ground power
cable to the power connection part on the back side of Electrosurgical Unit. If there is
no ground at Main Power, link an earth ground to the ground connection on the back
side of the product.
②
Turn on the Power Switch.
Check if Front Display indicates a number and then if REM alarm lamp flickers.
③
Connect Patient Return Plate Cable to the Patient Plate connection at the surgical unit.
- Single Patient Plate
Connect Single Return Plate to Patient Return plate Cable Clamp. Alarm lamp stops
flickering and the standby state starts. Attach Pad to a patient
’
s body.
- Double Patient Return Plate
Connect Double Patient Return
Plate to Patient Return plate Cable Clamp. Attach Pad
to a patient
’
body. Alarm lamp stops flickering and the standby state starts.
④
Select Cutting Mode (Cut, Blend) and set a wanted output (W) by using UP and
DOWN buttons.
⑤
Select Coagulation Mode (Contact) and set a wanted output (W) level by using UP
and DOWN buttons.
⑥
In case of using Twin Button Handle, connect it to the connection 1 for Cutting
and Coagulation Electrode.
⑦
In case of using Foot Switch, Connect Foot Switch to Monopolar Foot Switch
connector in the back of equipment.
Output is generated from the Connection 2 for Cutting and Coagulation Electrode.
⑧
Press Twin Button Handle Switch or Foot Switch