ZEUS-100/80 ZO-OM-008
REV : 08 9
ZERONE
Check the followings before using this product.
Warning
②
Use the power code provided by the equipment supplier or a other test-completed
product that has equal quality.
③
Don't use a extended code that has no ground connection.
④
Connection for potential equalization : Connection pins for potential equalization
cable between RF equipment and potential equalization bar in or room. The yellow-
green potential equalization cable supplied by the manufacturer must be used.
①
Don't place any equipment on Electrosurgical Unit, which can prevent this unit
from being cooled.
②
Do not leave any water container on the main system. If any water is spilled, wipe it
up immediately, and never run the system if water is soaked into it.
③
Electrosurgical Unit should be kept apart, as much as possible, from other
electronic equipment because the use of the unit can have negative effects on other
medical units that are using other electricity.
④
Check the linkage condition of equipment ground.
⑤
Impressed voltage (AC120V/AC230V) should be used in accordance with the rating.
Before providing power, check an operation voltage and power frequency marked in
the label on the back of the product. Inappropriate power could cause output
fluctuation and a direct fault in the product. To prevent over-standard voltage
fluctuation, it is recommended that Automatic Voltage Regulator (AVR) be used.
⑥
Before you start surgical operation, check if the power output is correctly adjusted
for your operation.
3. Application Notice
Electrosurgical Unit is supposed to generate high voltage and current for its
application. To avoid exposing a patient, an operator and the third parties
into a dangerous situation, operation should be carefully conducted, and
safety rules should be strictly kept.
●
About Patients
①
Eliminate or isolate all the patient
’
s metal materials and handle them with special
Interests
Check the followings before using this product.
Warning