5. Configuration
5.1
Feature
①
A DISPLAY window allows you to easily check output values, indicated as digits.
②
This provides necessary functions like Pure Cut, Blend, Contact Coagulation, Bipoar
Coagulation, under operation by one equipment
③
Holding the button for each mode for more than 3 seconds, the output ratio on
Display will be changed to 1W to be facilitated to shift from the low to high output.
④
Twin Button Handles provide remote-control function. (Select Cutting or Coagulation)
⑤
Foot Switch is available for Cutting and Coagulating.
⑥
Bipolar coagulation of by Foot Switch.
⑦
Microprocessor can straighten, stabilize the output.
⑧
Operations of Cutting, Coagulation and Bipolar Coagulation can be distinct from
each other by sound and indication lamp.
⑨
Each application mode (Cutting, Coagulation, Bipolar Coagulation) has a different
sound, which enables you to easily distinguish an operation type.
⑩
If the area between a patient and the pad is not appropriate, REM(Return Electrode
Monitoring) gives the alarm with a warning sound and stop an operation of the product
to prevent a burning incident.
⑪
The selected output ratio of Cut, Coagulation, and Bipolar Coagulation will displayed
by reenergizing the device after power OFF.
5.2
Safety Function
①
A fuse built in a power circuit prevent an over current from flowing through the
equipment.
②
When the plate attached to the patient is separated from the equipment, the red
alarm light begins flickering. Pressing the button of Twin Button Handle or pedal of
Foot Switch will stop the alarm sound and the equipment.
③
REM (Return Electrode Monitoring) monitors the size of contacting area between a
patient and the pad. If the size is inappropriate, it automatically blocks the high-
frequency current to minimize the danger of burning incidents.
④
To protect a patient, the case is fully grounded so that a leakage current can flow
into the earth.