About Restarting a Treatment
Treatment
48
BRZ-101-TUM-EN2-L
About Restarting a Treatment
A treatment can be interrupted by either the operator or the system. When you restart a treatment, the
treatment count on the card is not reduced further.
Each treatment can be restarted only once.
A treatment can be restarted if:
•
The operator interrupted the treatment during the first 10 minutes
•
The system interrupted the treatment during the first 10 minutes with one of the following Recoverable
Exceptions:
The coolant level is low.
Z403-YYY
Applicator control error.
Z408-YYY
Treatment quality error.
Z412-YYY
Potential loss of patient contact.
Z415-YYY
Interference detected.
Z426-YYY
•
And, the Restart timer interval of 60 minutes has not expired
Interrupt a Treatment
►
To interrupt a treatment:
1.
Press the Interrupt button.
The
Treatment Interrupted
screen is displayed.
The Restart Timer runs for up to 60 minutes, after which the treatment can no longer be restarted.
2.
Press the Next button to continue.
The Warming cue may be displayed for up to 2 minutes. When the applicator
cup is ready for the next treatment, the Next button is displayed.
Содержание CoolSculpting System
Страница 20: ...20 BRZ 101 TUM EN2 L This page intentionally left blank ...
Страница 36: ...Supplies System Overview 36 BRZ 101 TUM EN2 L This page intentionally left blank ...
Страница 52: ...Complete a Treatment Treatment 52 BRZ 101 TUM EN2 L This page intentionally left blank ...
Страница 68: ...Software Updates and Messages System Messages 68 BRZ 101 TUM EN2 L This page intentionally left blank ...
Страница 86: ...ZELTIQ Aesthetics Inc 4410 Rosewood Drive Pleasanton CA 94588 USA 925 474 2500 www coolsculpting com 10 2018 ...