System Tools
Settings Screen
User Manual
75
The
Settings
screen is displayed.
2.
To return to the System Tools screen, press the Previous button.
Calibration Screen
The system screen might require recalibration from time to time. If the screen does not respond accurately to
your touch, calibrate the screen.
►
To calibrate the screen:
1.
On the System Tools screen, press the
Settings
button.
2.
Press the
Calibration
button.
The
Calibration
screen is displayed.
3.
Use a cotton swab to press the cross-hatch.
The system records your touch and moves the cross-hatch to the next position.
4.
Press the cross-hatch in each position.
After you press the last setting, the system displays a message.
5.
To save your new settings, touch the screen within the time displayed in the message.
The new settings are saved and the
Settings
screen is displayed.