Treatment Sites
System Overview
4
BRZ-101-TUM-EN2-L
The use of other electronic medical devices on a patient who is undergoing a treatment might interfere with
the correct functioning of the system, possibly resulting in injury to the patient. Do not use other electronic
medical devices on a patient who is undergoing a treatment.
Treatment Sites
Observe the following warnings when treating the submental and submandibular areas:
Treatment Site
Warning
WARNING: Cold exposure to the
hypoglossal nerve may cause tongue
deviation following treatment of the
submental and submandibular areas.
Hypoglossal Nerve
WARNING: Cold exposure to the marginal
mandibular nerve may cause lower lip
weakness following treatment of the
submental and submandibular areas.
Marginal Mandibular Nerve
WARNING: Cold exposure to the
submandibular gland may cause
xerostomia, or decrease in saliva
production, following treatment of the
submental and submandibular areas.
Submandibular Gland
Table 1:
Submental and Submandibular Areas Treatment Warnings