Documentation set
7 Software description
VISULAS green
7.4 "Prepare" work step
000000-2096-101-DokS-GB-181218
59 / 116
Item
Icon/Name
Explanation
6
Pause
Set the interval between two laser pulses. The pause can be
adjusted using the rotary knob, tapping the two arrow keys or
with a circular motion on the displayed arc.
If the pause is shortened to below the minimum or the button is
activated
only a single laser pulse with the pre-set
duration will be emitted.
7
Aiming beam off / on
Aiming beam off / on. Use the [+] and [-] buttons to adjust the
brightness of the aiming beam.
8
Reset
Reset the laser spot counter to zero.
9
Energy density
Displays the energy density (laser fluence) of the laser beam at
the point to be treated. Energy density is a measure of thera-
peutic effect on the eye.
10
Laser pulse counter
Displays the number of triggered laser pulses.
11
Laser spot size
Displays the set laser spot size if no contact lens has been
selected.
If a retina therapy and contact lens have been selected, the laser
spot size on the retina is displayed (marked with "@ Retina").
If a glaucoma therapy and contact lens have been selected, the
laser spot size at the point of therapy is displayed (marked with
"@ Target").
12
Trigger
Displays the selected trigger.
Foot trigger has been selected.
Manual trigger has been selected.
13
Applicator
Displays the selected applicator.
Slit lamp has been selected as the applicator.
LIO VISULAS green laser indirect ophthalmoscope has been
selected as the applicator.
Laser endoprobe has been selected as the applicator.
Содержание VISULAS green
Страница 1: ...VISULAS green Documentation set...
Страница 20: ...Empty page for your notes...
Страница 42: ...Empty page for your notes...
Страница 48: ...Empty page for your notes...
Страница 92: ...Empty page for your notes...
Страница 110: ...Empty page for your notes...
Страница 112: ...Empty page for your notes...
Страница 114: ...Empty page for your notes...