Documentation set
7 Software description
VISULAS green
7.4 "Prepare" work step
000000-2096-101-DokS-GB-181218
61 / 116
Item
Icon/Name
Explanation
6
Aiming beam off / on
Aiming beam off / on. Use the [+] and [-] buttons to adjust the
brightness of the aiming beam.
7
Multi-spot sequence
Perform settings for multi-spot sequences (see Section Configure
the multi-spot sequence [
8
Titrate
Switch to Titrate mode. The laser power in Single-spot mode can
be optimized here.
9
Reset
Reset the laser spot counter to zero.
10
Energy density
Displays the energy density (laser fluence) of the laser beam at
the point to be treated. Energy density is a measure of thera-
peutic effect on the eye.
11
Laser pulse counter
Displays the number of triggered laser pulses.
12
Laser spot size
Displays the set laser spot size if no contact lens has been
selected.
If a retina therapy and contact lens have been selected, the laser
spot size on the retina is displayed (marked with "@ Retina").
If a glaucoma therapy and contact lens have been selected, the
laser spot size at the point of therapy is displayed (marked with
"@ Target").
13
Trigger
Displays the selected trigger.
Manual trigger
Manual trigger has been selected.
14
Applicator
Displays the selected applicator.
Slit lamp has been selected as the applicator.
Содержание VISULAS green
Страница 1: ...VISULAS green Documentation set...
Страница 20: ...Empty page for your notes...
Страница 42: ...Empty page for your notes...
Страница 48: ...Empty page for your notes...
Страница 92: ...Empty page for your notes...
Страница 110: ...Empty page for your notes...
Страница 112: ...Empty page for your notes...
Страница 114: ...Empty page for your notes...