DEVICE SAFETY
Laser TIRF 3
Safety Instructions
Carl Zeiss
02/2014
423683-7144-001
9
1.2.2
Installation Instructions
The Laser TIRF imaging system must be installed and commissioned by technicians trained in
laser safety and authorized by Carl Zeiss. The client is liable to pay the costs of services
rendered by the Carl Zeiss Microscopy GmbH.
To install the Laser TIRF imaging system, ensure a minimum distance of 0.5 m between the
wall and the rear of the system. This distance is necessary for adjustment and maintenance.
Do not set up the equipment near heat sources, such as heaters or direct sunlight. Do not
cover ventilation slots on the microscope system to avoid heat build-up.
The power plugs must be inserted in outlets with a grounding (earth) contact. The protective
capacity must not be rendered ineffective by using an extension cable without a ground wire.
Before connecting the power cables, verify that the available line voltage conforms to the
voltage specified on the type label of the laser bench tower or rack.
Always disconnect the power cable before opening the device and replacing fuses.
Only use fuses for the given rated current which comply with the specifications on the fuse
carriers and in this manual. The use of makeshift fuses and short-circuiting of fuse holders are
not permitted.
For reasons of laser safety, the C mount outputs of the microscope must be fitted either with
a camera or closed by a protective cap.
If any protective measures are no longer effective, the device must be taken out of service and
secured against inadvertent operation. Please contact a Carl Zeiss service agency or the
Carl Zeiss Microscopy Service department to repair the instrument.
Immersol 518 F
®
immersion oil irritates the skin. Avoid any contact with your skin, eyes or
clothing.
In case of actual skin contact, wash off with plenty of water and soap. After eye contact,
immediately rinse the eye with plenty of water for at least five minutes. If irritation persists,
consult a physician.
Proper disposal of Immersol 518 F
®
: Ensure that immersion oil does not enter surface water or
the sewage system.
Do not touch the hot lamp housings and adhere to a cooling time of about 15 minutes.
The device contains electronic components. At the end of its service life, it has to be disposed
of in compliance with the relevant national legislation.