OPERATION
Laser TIRF 3
Switching off
Carl Zeiss
02/2014
423683-7144-001
75
4.3
Switching off
The key switches on the Laser Bench Tower and Rack also serve as EMERGENCY STOP switches.
In case of danger, turn the key switch to the
OFF
or
0
position. This will switch off the lasers
immediately.
•
Turn off the laser (key switch on laser bench tower or rack to
OFF
). This will shut all laser lines down.
•
If required, close the ZEN software (blue edition) and shut down the control computer.
•
Switch of the signal distribution box and the TIRF Controller separately if necessary.
•
Turn off the entire system via the three switchable distributor sockets.
4.4
ZEN Software (blue edition)
The Slider Laser TIRF mot. is controlled via the ZEN software (blue edition).
For operating the software please see Section 4.7 and the
ZEN operating manual (blue edition).
4.5
Checking and Adjusting the Laser Beam Divergence and the TIRF Angle
TIRF divergence is adjusted using the
Focus
rotary button (Fig. 4-1/
1
) on the Slider Laser TIRF. The focus
of the laser beam can be adjusted to the pupil position of all usable objectives using this screw.
Every removal and re-installation of the Slider Laser TIRF must be followed by calibration.
Calibration of the Slider Laser TIRF man. is described in the following sections.
The Slider Laser TIRF mot. is controlled via the ZEN software (blue edition) using the Calibration Wizard.
Calibration of the Slider Laser TIRF mot. via software requires the camera adapter 60N-C 1.0x
for TIRF 3 with Bertrand optics.
4.5.1
Delivery Scope of the TIRF Adjustment Tool Kit
The following components are included in the TIRF adjustment tool kit (423684-9010-000) delivered:
−
Adjustment tool used to adjust laser beam divergence,
−
Adjustment tool used to adjust the TIRF angle,
−
Reflector module ACR P&C for reflected light with two ND filters 1 %,
−
2x Adapters M27x0.75 to W0.8; H=0 mm,
Содержание TIRF 3