Documentation set
3 Description of the device
SL 800
3.6 Control buttons and indicator lamps on the power supply unit
000000-2268-355-DokS-GB-300719
25 / 94
3.6.1 Configuration switches on power supply unit
Some functions of the device can be individually configured by
adjusting the four configuration switches on the power supply unit.
u
To set the configuration switches, use an object with a rounded
tip.
1
2
3
4
Figure 8: Configuration switches on power supply unit
No. Configuration Switch setting
Action
1
Not assigned
Not specified
Not specified
2
EcoMode
bottom (factory default):
automatic
The illumination of the device switches off automatically if
the user does not touch or move the joystick for approx.
3 minutes. The illumination of the device switches
automatically back to the preset brightness as soon as the
user touches the joystick again.
top: disabled
EcoMode is disabled.
3
QuickStop
bottom (factory default):
automatic
When the device changes to EcoMode, the QuickStop is
automatically enabled so that the instrument base is fixed
in the appropriate position.
The manual enabling and disabling of QuickStop remains
unaffected by the function described.
top: manual
QuickStop is not enabled when the device changes to
EcoMode. Quickstop can be enabled or disabled by
pressing the corresponding key.
Содержание SL 800
Страница 1: ...SL 800 Slit lamp Documentation set ...
Страница 6: ...Empty page for your notes ...
Страница 32: ...Empty page for your notes ...
Страница 34: ...Empty page for your notes ...
Страница 36: ...Empty page for your notes ...
Страница 44: ...Empty page for your notes ...
Страница 52: ...Empty page for your notes ...
Страница 64: ...Empty page for your notes ...
Страница 88: ...Empty page for your notes ...
Страница 90: ...Empty page for your notes ...
Страница 92: ...Empty page for your notes ...