INTRODUCTION
Axio Vert.A1
Notes on Instrument Safety
Carl Zeiss
05/2012
431030-7044-001
9
Particular attention should be paid to the warnings, cautions and notes listed below:
The manufacturer cannot assume any liability for any use other than the intended use. This
includes individual assemblies or single parts. The same applies to all and any service or repair
work performed by unauthorized personnel, which will also void any guarantee / warranty.
Be certain to connect the power plug to a grounded (earthed) power outlet only. The
protective effect must not be defeated by an extension cable that does not have a protective
grounding conductor.
If any protective measures are found to be no longer effective, put the instrument out of
operation and protect it against any inadvertent use. Contact Carl Zeiss Microscopy Service in
Germany (see Page 99) or the appropriate Carl Zeiss dealership abroad to get the appliance
repaired.
Axio Vert.A1 microscopes are not equipped with any special devices that offer protection
against caustic, potentially infectious,
toxic, radioactive or other health-impairing specimens.
When working with such specimens, be sure to observe all applicable legal requirements,
notably national accident prevention regulations.
Axio Vert.A1 inverted microscopes come with a power supply unit integrated into the stand.
This unit allows power supply voltages ranging from 100 V to 240 V ±10 % (50 Hz / 60 Hz)
to be used, without requiring any change of voltage on the microscope.
Before turning the power supply units for the HBO 50 / HBO 100 illuminators on, be sure to
check whether they are suitable for the existing supply voltage.
Always pull the power plug before opening the instrument and replacing fuses.
The equipment fuses used must match the intended current rating. Never employ any
makeshift fuses or short-circuit fuse holders.
Gas discharge lamps, such as HBO 100, emit ultraviolet radiation, which may cause burns on
eyes and on the skin. Therefore, be absolutely certain to avoid looking into the light and the
direct incidence of light on the skin. When microscoping, always use the protective devices
supplied with the instrument, such as specific attenuation filters. When heated, gas
discharge lamps are under high internal pressure. For this reason, be sure to replace them
only when they have cooled down. To do so, wear protective gloves and safety goggles.
When you utilize fluorescence filters, do not remove the heat protection filter, which provides
protection against the heat radiation of the microscope lamp. Since fluorescence filters are
heat-sensitive, removing the heat protection filter may adversely affect their functionality.
Clogged or covered ventilation slats may lead to heat build-up, which could damage the
instrument and, in extreme cases, cause a fire. Always keep ventilation slats clear and make
sure that no objects are put or can fall into them.